User Tools

Site Tools


mri-sequences

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
mri-sequences [2023/02/23 13:15] jonasmri-sequences [2024/02/19 14:52] (current) ari
Line 1: Line 1:
 ===== Structural ===== ===== Structural =====
-==== MPRAGE ====+==== T1-weighted ==== 
 +=== MPRAGE ===
 High resolution, T<sub>1</sub>-weighted, structural MRI High resolution, T<sub>1</sub>-weighted, structural MRI
  
Line 15: Line 16:
 | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 | | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 |
  
-==== For Incidental Findings ==== +====T2-weighted==== 
- +=== Turbo Spin Echo (TSE===
-=== TSE ===+
 Quick, high (in-plane) resolution, T<sub>2</sub>-weighted, structural MRI (for incidental findings) Quick, high (in-plane) resolution, T<sub>2</sub>-weighted, structural MRI (for incidental findings)
  
Line 34: Line 34:
 | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 26 | | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 26 |
  
-=== TSE_FLAIR ===+=== TSE FLAIR ===
 High (in-plane) resolution, T<sub>2</sub>-weighted, 'dark-fluid', structural MRI (for incidental findings) High (in-plane) resolution, T<sub>2</sub>-weighted, 'dark-fluid', structural MRI (for incidental findings)
  
Line 50: Line 50:
 | TE                       | 100 ms                                                        | | TE                       | 100 ms                                                        |
 | Flip angle               | 150 deg                                                        | Flip angle               | 150 deg                                                       
 +
 +===== Diffusion MRI =====
 +High resolution, T<sub>1</sub>-weighted, structural MRI
 +
 +^ Parameter                ^ Value                                                         ^
 +| Acquisition time         | 8 min 52 s                                                     |
 +| Type                     | 2D (Axial slices)                                          |
 +| b-value                  | 0, 1000                                     |
 +| Number of directions     | 64 (256 possible with a larger TA)                                        |
 +| FOV                      | 192 mm x 192 mm                                       |
 +| Matrix                   | 96 x 96                                     |
 +| Resolution               | 2 mm isotropic                                                |
 +| TR                       | 7800 ms                                                       |
 +| TE                       | 90 ms                                                       |
 +| Flip angle                                                                        | 
 +| Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 |
 +
  
 ===== BOLD fMRI ===== ===== BOLD fMRI =====
Line 217: Line 234:
 Other parameters are those of Fieldmapping_2iso. Other parameters are those of Fieldmapping_2iso.
  
----- +
-;#; +
-[[:useful_information|Top]] +
-;#;+
mri-sequences.1677158124.txt.gz · Last modified: 2023/02/23 13:15 by jonas

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki