User Tools

Site Tools


mri-sequences

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
mri-sequences [2024/02/08 12:20] – [For Incidental Findings] carlmri-sequences [2024/02/19 14:52] (current) ari
Line 1: Line 1:
 ===== Structural ===== ===== Structural =====
 ==== T1-weighted ==== ==== T1-weighted ====
 +=== MPRAGE ===
 High resolution, T<sub>1</sub>-weighted, structural MRI High resolution, T<sub>1</sub>-weighted, structural MRI
  
Line 15: Line 16:
 | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 | | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 |
  
-==== Diffusion MRI ====+====T2-weighted==== 
 +=== Turbo Spin Echo (TSE) === 
 +Quick, high (in-plane) resolution, T<sub>2</sub>-weighted, structural MRI (for incidental findings) 
 + 
 +^ Parameter                ^ Value                                                         ^ 
 +| Acquisition time         | 1 min 38 s                                                    | 
 +| Type                     | 2D                                                            | 
 +| Slices                   | 29                                                            | 
 +| Slice thickness          | 4 mm                                                          | 
 +| Slice gap                | 30% (1.2 mm)                                                  | 
 +| FOV                      | 220 mm x 220 mm                                               | 
 +| Matrix                   | 320 x 320                                                     | 
 +| Resolution               | 0.7 mm x 0.7 mm                                               | 
 +| TR                       | 5060 ms                                                       | 
 +| TE                       | 102 ms                                                        | 
 +| Flip angle               | 140 deg                                                       |  
 +| Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 26 | 
 + 
 +=== TSE FLAIR === 
 +High (in-plane) resolution, T<sub>2</sub>-weighted, 'dark-fluid', structural MRI (for incidental findings) 
 + 
 +^ Parameter                ^ Value                                                         ^ 
 +| Acquisition time         | 4 min 32 s                                                    | 
 +| Type                     | 2D                                                            | 
 +| Slices                   | 25                                                            | 
 +| Slice thickness          | 4 mm                                                          | 
 +| Slice gap                | 30% (1.2 mm)                                                  | 
 +| FOV                      | 220 mm x 220 mm                                               | 
 +| Matrix                   | 256 x 256                                                     | 
 +| Resolution               | 0.9 mm x 0.9 mm                                               | 
 +| TR                       | 9000 ms                                                       | 
 +| TI                       | 2500 ms                                                       | 
 +| TE                       | 100 ms                                                        | 
 +| Flip angle               | 150 deg                                                       |  
 + 
 +===== Diffusion MRI =====
 High resolution, T<sub>1</sub>-weighted, structural MRI High resolution, T<sub>1</sub>-weighted, structural MRI
  
 ^ Parameter                ^ Value                                                         ^ ^ Parameter                ^ Value                                                         ^
-| Acquisition time         min 32 s                                                     | +| Acquisition time         min 52 s                                                     | 
-| Type                     3D (sagittal slices)                                          | +| Type                     2D (Axial slices)                                          | 
-FOV (A-P x L-R x F-H   256 mm x 192 mm x 256 mm                                      +b-value                  | 0, 1000                                     | 
-| Matrix                   256 256 x 192 (rotated!)                                    +| Number of directions     | 64 (256 possible with a larger TA                                       | 
-| Resolution               mm isotropic                                                | +| FOV                      | 192 mm x 192 mm                                       
-| TR                       2250 ms                                                       +| Matrix                   96 96                                     
-| TI                       | 900 ms                                                        +| Resolution               mm isotropic                                                | 
-| TE                       2.98 ms                                                       | +| TR                       7800 ms                                                       | 
-| Flip angle               9 deg                                                         +| TE                       90 ms                                                       | 
 +| Flip angle                                                                        
 | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 | | Parallel imaging         | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 |
  
mri-sequences.1707394850.txt.gz · Last modified: 2024/02/08 12:20 by carl