===== Structural ===== ==== T1-weighted ==== === MPRAGE === High resolution, T1-weighted, structural MRI ^ Parameter ^ Value ^ | Acquisition time | 4 min 32 s | | Type | 3D (sagittal slices) | | FOV (A-P x L-R x F-H) | 256 mm x 192 mm x 256 mm | | Matrix | 256 x 256 x 192 (rotated!) | | Resolution | 1 mm isotropic | | TR | 2250 ms | | TI | 900 ms | | TE | 2.98 ms | | Flip angle | 9 deg | | Parallel imaging | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 | ====T2-weighted==== === Turbo Spin Echo (TSE) === Quick, high (in-plane) resolution, T2-weighted, structural MRI (for incidental findings) ^ Parameter ^ Value ^ | Acquisition time | 1 min 38 s | | Type | 2D | | Slices | 29 | | Slice thickness | 4 mm | | Slice gap | 30% (1.2 mm) | | FOV | 220 mm x 220 mm | | Matrix | 320 x 320 | | Resolution | 0.7 mm x 0.7 mm | | TR | 5060 ms | | TE | 102 ms | | Flip angle | 140 deg | | Parallel imaging | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 26 | === TSE FLAIR === High (in-plane) resolution, T2-weighted, 'dark-fluid', structural MRI (for incidental findings) ^ Parameter ^ Value ^ | Acquisition time | 4 min 32 s | | Type | 2D | | Slices | 25 | | Slice thickness | 4 mm | | Slice gap | 30% (1.2 mm) | | FOV | 220 mm x 220 mm | | Matrix | 256 x 256 | | Resolution | 0.9 mm x 0.9 mm | | TR | 9000 ms | | TI | 2500 ms | | TE | 100 ms | | Flip angle | 150 deg | ===== Diffusion MRI ===== High resolution, T1-weighted, structural MRI ^ Parameter ^ Value ^ | Acquisition time | 8 min 52 s | | Type | 2D (Axial slices) | | b-value | 0, 1000 | | Number of directions | 64 (256 possible with a larger TA) | | FOV | 192 mm x 192 mm | | Matrix | 96 x 96 | | Resolution | 2 mm isotropic | | TR | 7800 ms | | TE | 90 ms | | Flip angle | | | Parallel imaging | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 24 | ===== BOLD fMRI ===== T2*-weighted EPI sequences with BOLD contrast ==== Standard fMRI ==== Functional MRI with 3 mm isotropic spatial resolution === EPI_Standard === Standard fMRI, 2 s per volume ^ Parameter ^ Value ^ | Acquisition time | 2 s (per volume) | | Type | 2D | | Slices | 32 | | Slice thickness | 3 mm | | Slice gap | 25% (0.75 mm) | | Slice order | Descending (32 - 1) | | FOV | 192 mm x 192 mm | | Matrix | 64 x 64 | | Resolution | 3 mm x 3 mm | | TR | 2000 ms | | TE | 30 ms | | Flip angle | 78 deg | | Bandwidth | 2112 Hz/Px | | Echo spacing | 0.54 ms | === EPI_Standard_1s === Standard fMRI, 1 s per volume with reduced brain coverage ^ Parameter ^ Value ^ | Acquisition time | 1 s (per volume) | | Slices | 16 | | Slice order | Descending (16 - 1) | | TR | 1000 ms | | Bandwidth | 2232 Hz/Px | | Echo spacing | 0.51 ms | Other parameters are those of EPI_Standard. ==== High Resolution fMRI ==== Functional MRI with 2 mm isotropic or higher spatial resolution === EPI_HighRes_2iso === fMRI with 2 mm isotropic spatial resolution, 2 s per volume. Keep in mind that the thinner slices leads to reduced brain coverage! ^ Parameter ^ Value ^ | Acquisition time | 2 s (per volume) | | Type | 2D | | Slices | 32 | | Slice thickness | 2 mm | | Slice gap | 25% (0.5 mm) | | Slice order | Descending (32 - 1) | | FOV | 192 mm x 192 mm | | Matrix | 96 x 96 | | Resolution | 2 mm x 2 mm | | TR | 2000 ms | | TE | 30 ms | | Flip angle | 78 deg | | Bandwidth | 1628 Hz/Px | | Echo spacing | 0.7 ms | | Parallel imaging | GRAPPA \\ Acceleration factor PE: 2 \\ Reference lines PE: 48 | === EPI_HighRes_2iso_WholeBrain === fMRI with 2 mm isotropic spatial resolution, 40 slices for a whole brain coverage. Keep in mind that acquiring more slices leads to longer acquisition time! ^ Parameter ^ Value ^ | Acquisition time | 2.5 s (per volume) | | Slices | 40 | | TR | 2500 ms | Other parameters are those of EPI_HighRes_2iso. === EPI_HighRes_2iso_LargerFOV === fMRI with 2 mm isotropic spatial resolution, with a FOV of 205 mm for patricipants with larger head. Keep in mind that acquiring slices with larger FOV leads to slightly fewer slices thus slightly more reduced brain coverage! ^ Parameter ^ Value ^ | Slices | 30 | | FOV | 205 mm x 205 mm | | Matrix | 102 x 102 | | Bandwidth | 1400 Hz/Px | | Echo spacing | 0.8 ms | Other parameters are those of EPI_HighRes_2iso. === EPI_HighRes_1.5x1.5x2mm === fMRI with 1.5 mm in-plane (highest achievable) spatial resolution, 2 s per volume. Keep in mind that acquiring slices with higher resolution leads to slightly fewer slices thus slightly more reduced brain coverage! ^ Parameter ^ Value ^ | Acquisition time | 2 s (per volume) | | Slices | 28 | | Matrix | 128 x 128 | | Resolution | 1.5 mm x 1.5 mm | | Bandwidth | 1502 Hz/Px | | Echo spacing | 0.75 ms | Other parameters are those of EPI_HighRes_2iso. ===== Fieldmap ===== Magnetic field mapping sequences required to * undistort EPI data (e.g. fMRI, DWI) * measure functional sensitivity :!: It has to be acquired with the same geometry (FOV, resolution, alignment, slice centre) as the EPI data! :!: ==== Standard Fieldmap ==== Fieldmap with 3 mm isotropic spatial resolution. It matches with Standard fMRI sequences. === Fieldmapping === Dual-echo fieldmap ^ Parameter ^ Value ^ | Acquisition time | 54 s | | Type | 2D | | Slices | 32 | | Slice thickness | 3 mm | | Slice gap | 25% (0.75 mm) | | FOV | 192 mm x 192 mm | | Matrix | 64 x 64 | | Resolution | 3 mm x 3 mm | | TR | 400 ms | | TE | 5.19 ms, 7.65 ms | | Flip angle | 60 deg | ==== High Resolution Fieldmaps ==== Fieldmaps 2 mm isotropic or higher spatial resolution. They match with High resolution fMRI sequences. === Fieldmapping_2iso === High resolution dual-echo fieldmap ^ Parameter ^ Value ^ | Acquisition time | 1 m 19 s | | Type | 2D | | Slices | 32 | | Slice thickness | 2 mm | | Slice gap | 25% (0.5 mm) | | FOV | 192 mm x 192 mm | | Matrix | 96 x 96 | | Resolution | 2 mm x 2 mm | | TR | 400 ms | | TE | 5.19 ms, 7.65 ms | | Flip angle | 60 deg | === Fieldmapping_2iso_WholeBrain === High resolution dual-echo fieldmap with whole brain coverage ^ Parameter ^ Value ^ | Acquisition time | 1 m 26 s | | Slices | 40 | | TR | 437 ms | Other parameters are those of Fieldmapping_2iso. === Fieldmapping_2iso_LargerFOV === High resolution dual-echo fieldmap with larger FOV ^ Parameter ^ Value ^ | Acquisition time | 1 m 24 s | | Slices | 30 | | FOV | 205 mm x 205 mm | | Matrix | 102 x 102 | Other parameters are those of Fieldmapping_2iso. === Fieldmapping_1.5x1.5x2mm === High resolution dual-echo fieldmap with our highest achievable in-plane resolution ^ Parameter ^ Value ^ | Acquisition time | 1 m 45 s | | Slices | 32 | | Matrix | 128 x 128 | | Resolution | 1.5 mm x 1.5 mm | | Other parameters are those of Fieldmapping_2iso.